Capnostream™ 20p bedside capnography monitor

The Capnostream™ 20p patient monitor with Microstream™ technology provides a complete measurement of a patient’s oxygenation and ventilation status.

Description

An accurate and continuous view of ventilation

The Capnostream™ 20p patient monitor is built on a legacy of proven performance. For nearly two decades, clinicians have relied on Microstream™ technology for an accurate, reliable view of ventilation adequacy on intubated and non-intubated patients, from neonate to adult.1

The Capnostream™ 20p monitor, with the apnea-sat alert feature, alerts caregivers to repetitive patterns of apneas and oxygen desaturation, revealing a more complete picture of a patient’s respiratory status.

Features

  • The dual parameter monitor supports current standards of care by providing etCO2 and SpO2 measurements.

  • The monitor is compatible with central monitoring systems and interfaces with nurse call systems and electronic medical records.

  • Smart Alarm Management™ integrated algorithms2,3 include Smart Alarm Management™ integrated algorithms, a suite of algorithms proven to reduce alarms and simplify the use of patient monitoring to enhance patient safety and improve clinical efficiency.4

  • Smart Breath Detection™ (SBD) screens out low amplitude "non-breath" etCO2 excursions like talking, crying, or snoring.

  • Smart Alarm for Respiratory Analysis™ (SARA) reduces 53% of clinically insignificant, unactionable nuisance alarms.5

  • Nellcor™ SatSeconds alerts clinicians to clinically significant changes in pulse oximetry.

  • Integrated Pulmonary Index™ algorithm (IPI) presents one value demonstrating real-time respiratory status based on etCO2, RR, PR, and SpO2.†

  • Apnea-Sat Alert™ algorithm (ASA)‡ tracks and reports apneas per hour and oxygen desaturation index and provides the data in trend reports.

Specifications

Power supply

Input voltage

100-240VAC, 50/60Hz

Fuses

Two F3.15A 250 volt

Input power

90 VA

Battery

Battery type

14.4V, 4.5Ah lithium-ion

Battery operation

2.5 hours (without thermal recorder)

Battery charging time

100% in 12 hours

Controls

Front panel

1 switch for monitor on/off control
4 specific function keys
1 optical encoder with switch

Display

Screen

162 mm (6.4 in) color TFT display
Pixel pitch: 0.204 (horizontal) x 0.204 (vertical) mm
Active display area: 130.56 (horizontal) x 97.92 (vertical) mm (5.14 x 3.86 in)
Resolution: 640 x 480 pixels
Viewing angle: (vertical) 110 degrees
Viewing angle: (horizontal) 140 degrees

Trace speed

3.0, 6.3, 12.5, and 25 mm/second

Waveform sampling rate

75.7 samples/second for SpO2 (fixed)
20 samples/second for capnography (fixed)

Trend storage

8640-point storage
12 hours at 5 seconds resoltuion
24 hours at 10 seconds resolution
72 hours at 30 seconds resolution

Trend display

Graphical display: 2-hour, 6-hour, 12-hour views
Tabular display intervals: 60 minutes, 15 minutes, 3 minutes, and minimum resolution
(minimum resolution settable to 5, 10 or 30 seconds)

Microstream™ capnography

CO2 units

mmHg or kPa or Vol%

CO2, etCO2 FiCO2 range

0–150 mmHg

CO2, waveform resolution

0.1 mmHg

EtCO2, FiCO2 resolution

1 mmHg

CO2 accuracy

0–38 mmHg: ±2 mmHg
39–150 mmHg: ± (5% of reading + 0.08 for every 1 mmHg above 38 mmHg)

Respiration rate range

0–150 bpm

Respiration rate accuracy

0–70 bpm: ±1 bpm
71–120 bpm: ±2 bpm
121–150 bpm: ±3 bpm

CO2 alarms

No breath, etCO2 High, etCO2 Low, RR High, RR Low, Integrated Pulmonary Index™ algorithm (IPI).  IPI also requires pulse oximetry information

Flow rate

50 (42.5 ≤ flow ≤ 65) mL/minute, flow measured by volume

Waveform sampling

20 samples/second

Response time

2.95 seconds (typical); with use with sampling lines with long tubing, ~5.0 seconds

Initialization time

40 seconds (typical)

Calibration interval

Initially calibrate after 1,200 operating hours, then once a year or after 4,000 operating hours, whichever comes first

Nellcor™ pulse oximetry

SpO2 measurement range

1 to 100%

SpO2 accuracy

Neonatal to adult

Saturation (% SpO2 ±1 SD)
Without motion: 70 to 100% +/- 2 digits
With motion: 70-100% +/- 3 digits

Low perfusion

70 to 100% ±2 digits
20 to 250 bpm ±3 digits

Alarms

Adjustable alarm limits
SpO2 High, SpO2 Low, PR High, PR Low

Nellcor™ SatSeconds alarm management range

10, 25, 50, 100

Data output and connectivity

Data output

USB storage to flash memory devices
Analog output for use in sleep labs and other laboratory enviornments
Serial port (RS-232) for data transfer

Connectivity options

Vital Sync™ virtual patient monitoring platform connectivity option
Profox™* respiratory oximetry software
Nurse call
Nuvon VEGA™* system
iSirona™* system
Capsule™* system
Cerner™* system
Epic™* system

Hours